Specimen Collection and Preparation: Urine Assay: The urine specimen should be collected in a clean and dry container. Urine collected at anytime of the day may be used. Urine specimens exhibiting visible precipitates should be centrifuged, filtered, or allowed to settle to obtain a clear specimen for testing.
Specimen Storage: Urine specimens may be stored at 2-8°C for up to 48 hours prior to testing. For prolonged storage, specimens may be frozen and stored below -20°C. Frozen specimens should be thawed and mixed well before testing.
Directions for Use: Allow the test, urine specimen, and/or controls to reach room temperature (15-30°C) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test panel from the sealed pouch and use it within one hour.
2. Remove the cap.
3. With the arrow pointing toward the urine specimen, immerse the test panel vertically in the urine specimen for at least 10 to 15 seconds. Immerse the dipstick to at least the level of the wavy lines, but not above the arrow on the test panel.
4. Replace the cap and place the test panel on a non-absorbent flat surface.
5. Start the timer and wait for the colored line(s) to appear.
6. The drug strip result should be read at 5 minutes. Do not interpret the result after 10 minutes.
Interpretation of Results: NEGATIVE: *A colored line appears in the control region (C) and another colored line appears in the test region (T). This negative result means that the concentrations in the urine sample are below the designated cut-off levels for a particular drug tested.
*NOTE: The shade of the colored lines(s) in the test region (T) may vary. The result should be considered negative whenever there is even a faint line.
POSITIVE: A colored line appears in the control region (C) and no line appears in the test region (T). The positive result means that the drug concentration in the urine sample is greater than the designated cut-off for a specific drug.
INVALID: No line appears in the control region (C). Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Read the directions again and repeat the test with a new test. If the result is still invalid, contact the manufacturer. (See figure.)
Click on icon to see table/diagram/image
Quality Control: A procedural control is included in the test. A line appearing in the control region (C) is considered an internal procedural control. It confirms sufficient specimen volume, adequate membrane wicking, and correct procedural technique.
Control standards are not supplied with this kit. However, it is recommended that positive and negative controls be tested as good laboratory practice to confirm the test procedure and to verify proper test performance.
Limitations: The Multi-Drug Rapid Test Panel provides only a qualitative, preliminary result. A secondary analytical method must be used to obtain a confirmed result. Gas Chromatograph/Mass Spectrometry (GCIMS) is the preferred confirmatory method.
There is a possibility that technical or procedural errors, as well as interfering substances inthe urine specimen, may cause erroneous results.
Adulterants, such as bleach and/or alum, in urine specimens may produce erroneous results regardless of the analytical method used. If adulteration is suspected, the test should be repeated with another urine specimen.
A positive result does not indicate level or intoxication, administration route, or concentrationin urine.
A negative result may not necessarily indicate drug-free urine. Negative results can be obtained when drug is present but below the cut-off level of the test.
This test does not distinguish between drugs of abuse and certain medications.
A positive test result may be obtained from certain foods or food supplements.